Disclaimer: This article is for informational purposes. Always follow the official documentation provided by Matachana/Getinge for your specific autoclave model. Improper servicing can lead to severe injury or equipment damage.

The manual provides the technician passcode to access the service menu on the PLC touchscreen. (Usually a combination like Menu -> Service -> Code 1234 , but varies by model). matachana autoclave service manual

Navigate to the "Leak Test" function as per Section 4.2. The autoclave will pull a vacuum to 100 mbar and hold it for 10 minutes. The manual states the acceptable rise is <1.3 mbar/min. Disclaimer: This article is for informational purposes